Phosphorylation is a reversible posttranslational protein modification which plays a pivotal role in intracellular signaling. Despite extensive efforts, phosphorylation site mapping of pro- teomes is still incomplete motivating the exploration of alternative methods that complement existing workflows. In this study, we compared tandem mass spectrometry (MS/MS) on matrix assisted laser desorption/ionization time-of-flight (MALDI-TOF) and nano-electrospray ioniza- tion (nESI) Orbitrap instruments with respect to their ability to identify phosphopeptides from complex proteome digests. Phosphopeptides were enriched from tryptic digests of cell lines using Fe-IMAC column chromatography and subjected to LC-MS/MS analysis. We found that the two analytical workflows exhibited considerable orthogonality. For instance, MALDI-TOF MS/MS favored the identification of phosphopeptides encompassing clear motif signatures for acidic residue directed kinases. The extent of orthogonality of the two LC-MS/MS systems was comparable to that of using alternative proteases such as Asp-N, Arg-C, chymotrypsin, Glu-C and Lys-C on just one LC-MS/MS instrument. Notably, MALDI-TOF MS/MS identified an un- expectedly high number and percentage of phosphotyrosine sites (20% of all sites), possibly as a direct consequence of more efficient ionization. The data clearly show that LC-MALDI MS/MS can be a useful complement to LC-nESI MS/MS for phosphoproteome mapping and particularly so for acidic and phosphotyrosine containing peptides.
NEWS - ARCHIVE
MALDI‐TOF and nESI Orbitrap MS/MS identify orthogonal parts of the phosphoproteome